PROTECTION ISOLATION TRANSFORMERS FOR MEDICAL PURPOSES

STN EN 60601-1, IEC 61558-2-15

transformátor                 transformátor

 

TYPE CAPACITY
[VA]
CORE DIMENSIONS
VOLTAGE
[V]
WEIGHT
[kg]
TR-9415/4-3k MED 1000 EI 192/70 230/230 20
TR-9415/4-3k MED 1500 EI 192/100 230/230 25
TR-9415/4-3k MED 2000 EI 192/130 230/230 35
TR-9415/4-3k MED 2500 UI 180/60 230/230 40
TR-9415/4-3k MED 3000 UI 180/80 230/230 47
TR-9415/4-4k MED 4000 UI 180/100 230/230 55
TR-9415/4-4k MED 5000 UI 210/80 230/230 73
TR-9415/4-6,3k MED 6300 UI 210/130 230/230 85
TR-9415/4-6,3k MED 8000 UI 210/130 230/230 100
 
SHORT-CIRCUIT RESISTANCE: NOT RESISTANT
IMPREGNATION: Electroinsulating baking varnish
INPUT VOLTAGE: 230V 50Hz
PROTECTION: IP00
ELECTRIC RESISTANCE: 4700V
THERMAL CLASS: E
 
Notes:
 
  • the manufacturer recommends protection by a circuit breaker with “D” characteristic as mentioned on the plate
  • the transformer’s secondary side is beyond the central output for the isolation condition monitoring
  • to control, a thermal sensor to 1.5 A is installed
 

 

DIMENSIONS

rozmery transformátora

CAPACITY [VA]
A B C D E
1000 192 130 172 127 100
1500 192 176 172 127 120
2000 192 207 172 127 140
2500 250 140 360 143 100
3000 250 160 360 143 123
4000 250 230 360 143 150
5000 280 210 410 210 130
6300 280 180 410 210 180
8000 280 180 410 210 180